5-30 | IPS-4-UV Analyzer-GPDiv2
Condition/Alarm Name
Description & Suggested Corrective Action
Manifold pressure
Aspirator pressure
The
se alarm conditions indicate that the pressure within these
components exceeds its full-scale range by more than 10 psia (t
ypical;
set point and pressure unit may vary).
Corrective Action:
•
The parameters have exceeded the normal operating range. From
the User Interface Panel, check that the full-scale range is correct for
the current sample. If
Manifold
or
Aspirator pressure
alarms persist
during normal operation, contact the factory. A range change and
recalibration may be required.
Internal error
This alarm condition indicates that the analyzer has detected a serious
malfunction.
Corrective Action:
Take appropriate safety precautions, open the Electronics Enclosure
door and:
•
Check all of the ribbon cables between boards for proper connections
and inspect them for damage (cuts, nicks, burn marks, etc.). For
cable connection locations, refer to the Electronics Enclosure Wiring
Diagram in the Appendix.
•
Contact AMETEK for assistance.
Failed communications to the Analog board
This alarm condition indicates that the
MCU
board cannot establish
reliable communications with the
Analog
board.
Corrective Action:
Take appropriate safety precautions, open the Electronics Enclosure
door and:
•
Check the ribbon cable (
Cable 3
) between the
MCU
board (
P3
) and
Analog
board (
J101
) – both boards mounted on the Electronics
Enclosure door – for proper connections and inspect it for damage
(cuts, nicks, burn marks, etc.). For cable connection locations, refer to
the Electronics Enclosure Wiring Diagram in the Appendix.
•
One or both boards may need to be replaced.
Contact AMETEK for
assistance.
Failed communications to the Relay board
This alarm condition indicates that the
MCU
board cannot establish
reliable communications with the
Relay
board.
Corrective Action:
Take appropriate safety precautions, open the Electronics Enclosure
door and:
•
Check the fl at cable (
Cable 3
) between the
MCU
board (
P3
, on
the Electronics Enclosure door) and
Relay
board (
J101
, on the
Electronics Enclosure backpan) for proper connections and inspect
it for damage (cuts, nicks, burn marks, etc.). For cable connection
locations, refer to the Electronics Enclosure Wiring Diagram in the
Appendix.
•
One or both boards may need to be replaced.
Contact AMETEK for
assistance.
Содержание IPS-4-UV
Страница 15: ...Contents xv...
Страница 16: ...xvi IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 24: ...1 8 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 66: ...3 36 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 138: ...5 32 IPS 4 UV Analyzer GPDiv2 Figure 5 4 Over Temperature alarm Reset Switch S501 Relay board 100 1942...
Страница 144: ...6 6 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 180: ...A 2 IPS 4 UV Analyzer GPDiv2 Analyzer Light Path Schematic...
Страница 183: ...Appendix Drawings A 5 Detector Interface Board 100 1937...
Страница 184: ...A 6 IPS 4 UV Analyzer GPDiv2 Detector Board 100 1955...
Страница 185: ...Appendix Drawings A 7 Display Interface Board 100 1978...
Страница 186: ...A 8 IPS 4 UV Analyzer GPDiv2 Customer Connection Board 100 1971...
Страница 187: ...Appendix Drawings A 9 Relay Board 100 1942...
Страница 188: ...A 10 IPS 4 UV Analyzer GPDiv2 Analog Board 100 1922...
Страница 189: ...Appendix Drawings A 11 MCU Board 100 1914...
Страница 190: ...A 12 IPS 4 UV Analyzer GPDiv2 Xenon Lamp Board 100 2061...
Страница 191: ...Appendix Drawings A 13 Electronics Enclosure Wiring Diagram Sheet 1 of 3 WX 102877...
Страница 192: ...A 14 IPS 4 UV Analyzer GPDiv2 Electronics Enclosure Wiring Diagram Sheet 2 of 3 WX 102877...
Страница 193: ...Appendix Drawings A 15 Electronics Enclosure Wiring Diagram Sheet 3 of 3 WX 102877...
Страница 194: ...A 16 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 196: ...S 2 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 200: ...G 4 IPS 4 UV Analyzer GPDiv2 g 1035 17 32 Not used 1036 Not used 1037 Not used 1038 Not used 1039 Not used...