Installation & Start-Up | 3-29
Manually Zeroing the Analyzer Sample System
The Zero function forces the analyzer to purge the sample system with a
Zero fluid to remove any residual contaminants in the sample system.
For heated systems, all temperature zones must be at their operating set
points before performing a Zero. If necessary, view the temperature zones
to verify this.
Pressing
CANCEL
at any time during this procedure will abort the
procedure and return the software to the
HOME
screen.
To manually zero the analyzer:
1. Ensure that the Zero fluid cylinder (Instrument Air or Nitrogen) is
connected and turned on.
2. Press
the
SETUP
soft key to display the
SETUP
menu. Select the
Cali-
bration Control
sub-menu and press enter. Scroll down to the Cali-
bration sub-menu and press enter to display the
Calibration
screen.
3. Use
the
DOWN
arrow key to select
Start zero calibration
and press
EN-
TER
to purge the sample system with Zero fluid. The message “
Please
wait ...
” will appear while the analyzer performs the calibration func-
tion.
Run the zero for 10 minutes.
Typically, the IPS-4 Analyzer controls the zero flow rate; no adjust-
ments are required.
If the customer provides the sample system zero, run the manual zero
at a flow rate of 2.5 L/minute (0.08 scfm).
4.
From the
Calibration
screen, select
Start measure
and press
ENTER
to
return the analyzer to Measure mode.
5. Press
the
HOME
soft key to return to the
HOME
screen.
6. Turn off the Zero fluid.
The zero is complete and the analyzer is now on-line.
!
CAUTION
NOTE
Содержание IPS-4-UV
Страница 15: ...Contents xv...
Страница 16: ...xvi IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 24: ...1 8 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 66: ...3 36 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 138: ...5 32 IPS 4 UV Analyzer GPDiv2 Figure 5 4 Over Temperature alarm Reset Switch S501 Relay board 100 1942...
Страница 144: ...6 6 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 180: ...A 2 IPS 4 UV Analyzer GPDiv2 Analyzer Light Path Schematic...
Страница 183: ...Appendix Drawings A 5 Detector Interface Board 100 1937...
Страница 184: ...A 6 IPS 4 UV Analyzer GPDiv2 Detector Board 100 1955...
Страница 185: ...Appendix Drawings A 7 Display Interface Board 100 1978...
Страница 186: ...A 8 IPS 4 UV Analyzer GPDiv2 Customer Connection Board 100 1971...
Страница 187: ...Appendix Drawings A 9 Relay Board 100 1942...
Страница 188: ...A 10 IPS 4 UV Analyzer GPDiv2 Analog Board 100 1922...
Страница 189: ...Appendix Drawings A 11 MCU Board 100 1914...
Страница 190: ...A 12 IPS 4 UV Analyzer GPDiv2 Xenon Lamp Board 100 2061...
Страница 191: ...Appendix Drawings A 13 Electronics Enclosure Wiring Diagram Sheet 1 of 3 WX 102877...
Страница 192: ...A 14 IPS 4 UV Analyzer GPDiv2 Electronics Enclosure Wiring Diagram Sheet 2 of 3 WX 102877...
Страница 193: ...Appendix Drawings A 15 Electronics Enclosure Wiring Diagram Sheet 3 of 3 WX 102877...
Страница 194: ...A 16 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 196: ...S 2 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 200: ...G 4 IPS 4 UV Analyzer GPDiv2 g 1035 17 32 Not used 1036 Not used 1037 Not used 1038 Not used 1039 Not used...