5-20 | IPS-4-UV Analyzer-GPDiv2
Replacing the Xenon Flash Lamp Assembly
To replace the Xenon Flash Lamp Assembly:
1. Manually zero the analyzer for 10 minutes.
2. After the zero is complete, close the Sample Line and then the Vent
Line Shut Off valves (in the Oven/Sample System Enclosure) to isolate
the analyzer from the sample stream.
3. Power down the analyzer:
Hazardous Location Applications:
If an external, flame-proof disconnect switch (breaker) has been
installed, open the breaker. If no external breaker has been installed,
follow the power down procedure for “General Purpose (GP) Applica-
tions,” below.
Before proceeding, test the area around the analyzer for flammable
gases. If an explosive gas atmosphere is present, do not power down
the analyzer or any alternate power sources that supply power to the
analyzer components. Proceed only when the area is found to be safe.
General Purpose (GP) applications:
Open the Electronics Enclosure door and disconnect AC power from
the analyzer and its temperature-control zone circuitry by opening the
Main AC Power
,
DC Power
,
Bench Heater
, and
Oven Heater
(if used)
fuses. Refer to the Fuse Legend at the back of the Electronics Enclo-
sure for the locations of these fuses.
For all applications:
Wait 2 minutes to allow the high-voltage capacitors to discharge.
For all applications:
If heated Sample and Vent Lines are used, disconnect power to them.
4. Disconnect
the
DC Power Cable
connector (from the Lamp Housing)
at
J2
on the
Xenon Lamp
board located near the top of the Electronics
Enclosure (see Figure 5-1 for location).
!
WARNING
!
WARNING
NOTE
Содержание IPS-4-UV
Страница 15: ...Contents xv...
Страница 16: ...xvi IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 24: ...1 8 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 66: ...3 36 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 138: ...5 32 IPS 4 UV Analyzer GPDiv2 Figure 5 4 Over Temperature alarm Reset Switch S501 Relay board 100 1942...
Страница 144: ...6 6 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 180: ...A 2 IPS 4 UV Analyzer GPDiv2 Analyzer Light Path Schematic...
Страница 183: ...Appendix Drawings A 5 Detector Interface Board 100 1937...
Страница 184: ...A 6 IPS 4 UV Analyzer GPDiv2 Detector Board 100 1955...
Страница 185: ...Appendix Drawings A 7 Display Interface Board 100 1978...
Страница 186: ...A 8 IPS 4 UV Analyzer GPDiv2 Customer Connection Board 100 1971...
Страница 187: ...Appendix Drawings A 9 Relay Board 100 1942...
Страница 188: ...A 10 IPS 4 UV Analyzer GPDiv2 Analog Board 100 1922...
Страница 189: ...Appendix Drawings A 11 MCU Board 100 1914...
Страница 190: ...A 12 IPS 4 UV Analyzer GPDiv2 Xenon Lamp Board 100 2061...
Страница 191: ...Appendix Drawings A 13 Electronics Enclosure Wiring Diagram Sheet 1 of 3 WX 102877...
Страница 192: ...A 14 IPS 4 UV Analyzer GPDiv2 Electronics Enclosure Wiring Diagram Sheet 2 of 3 WX 102877...
Страница 193: ...Appendix Drawings A 15 Electronics Enclosure Wiring Diagram Sheet 3 of 3 WX 102877...
Страница 194: ...A 16 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 196: ...S 2 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 200: ...G 4 IPS 4 UV Analyzer GPDiv2 g 1035 17 32 Not used 1036 Not used 1037 Not used 1038 Not used 1039 Not used...