Supplemental Information
|
G-1
Appendix G
System alarm and warning bit map
When System Status (parameter 138, 16-bit integer, modbus register 126) contains none-zero value, th
bit map of System Alarms (parameter 134, 135, 136; 32-bit integer, modbus register 118, 119, 120,
121, 122, and 123) contains system alarms as defined:
Modbus
register
Bit Description
Type
Condition
118
1 (User-defined message)
Alarm
These alarms are defined by users
and configured in the state machine
and/or the menu XML configuration
files
118
2
(User-defined message)
Alarm
118
3
(User-defined message)
Alarm
118
4
(User-defined message)
Alarm
118
5
(User-defined message)
Alarm
118
6
(User-defined message)
Alarm
118
7
(User-defined message)
Alarm
118
8
(User-defined message)
Alarm
118
9 Heater temperature high limit
alarm
Alarm
Heater temperature (292) is outside
of a band set by parameter 287
around the setpoint (293)
118
10 Heater temperature low limit
alarm
Alarm
118
11 Bench temperature high limit
alarm
Alarm
Bench temperature (283) is outside
of a band set by parameter 278
around the setpoint (284)
118
12 Bench temperature low limit
alarm
Alarm
118
13 Lamp temperature high limit
alarm
Alarm
Lamp temperature (302) is outside of
a band set by parameter 297 around
the setpoint (303)
118
14 Lamp temperature low limit
alarm
Alarm
118
15 Cell temperature high limit
alarm
Alarm
Cell temperature (311) is outside of a
band set by parameter 306 around
the setpoint (312)
118
16 Cell temperature low limit
alarm
Alarm
118
17 Pressure tranducer 1 high
limit alarm
Alarm
Pressure tranducer 1 (338) is higher
than high limit (366)
118
18 Pressure tranducer 1 low limit
alarm
Alarm
Pressure tranducer 1 (338) is lower
than low limit (367)
118
19 Pressure tranducer 2 high
limit alarm
Alarm
Pressure tranducer 2 (341) is higher
than high limit (368)
118
20 Pressure tranducer 2 low limit
alarm
Alarm
Pressure tranducer 2 (341) is lower
than low limit (369)
118
21 Manifold pressure high limit
alarm
Alarm
Manifold pressure (344) is higher
than high limit (370)
118
22 Manifold pressure low limit
alarm
Alarm
Manifold pressure (344) is lower than
low limit (371)
Содержание IPS-4-UV
Страница 15: ...Contents xv...
Страница 16: ...xvi IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 24: ...1 8 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 66: ...3 36 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 138: ...5 32 IPS 4 UV Analyzer GPDiv2 Figure 5 4 Over Temperature alarm Reset Switch S501 Relay board 100 1942...
Страница 144: ...6 6 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 180: ...A 2 IPS 4 UV Analyzer GPDiv2 Analyzer Light Path Schematic...
Страница 183: ...Appendix Drawings A 5 Detector Interface Board 100 1937...
Страница 184: ...A 6 IPS 4 UV Analyzer GPDiv2 Detector Board 100 1955...
Страница 185: ...Appendix Drawings A 7 Display Interface Board 100 1978...
Страница 186: ...A 8 IPS 4 UV Analyzer GPDiv2 Customer Connection Board 100 1971...
Страница 187: ...Appendix Drawings A 9 Relay Board 100 1942...
Страница 188: ...A 10 IPS 4 UV Analyzer GPDiv2 Analog Board 100 1922...
Страница 189: ...Appendix Drawings A 11 MCU Board 100 1914...
Страница 190: ...A 12 IPS 4 UV Analyzer GPDiv2 Xenon Lamp Board 100 2061...
Страница 191: ...Appendix Drawings A 13 Electronics Enclosure Wiring Diagram Sheet 1 of 3 WX 102877...
Страница 192: ...A 14 IPS 4 UV Analyzer GPDiv2 Electronics Enclosure Wiring Diagram Sheet 2 of 3 WX 102877...
Страница 193: ...Appendix Drawings A 15 Electronics Enclosure Wiring Diagram Sheet 3 of 3 WX 102877...
Страница 194: ...A 16 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 196: ...S 2 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 200: ...G 4 IPS 4 UV Analyzer GPDiv2 g 1035 17 32 Not used 1036 Not used 1037 Not used 1038 Not used 1039 Not used...