Controller / User Interface | 4-35
Figure 4-36.
Ethernet Network
screen.
Ethernet Enable
Enabled
DHCP
Enable
Disabled
IP
Address
##.##.###.#
IP Subnet Mask
###.###.#.#
Gateway
##.##.#.#
MAC
address
##-##-##-##-##-##
HOME ALARMS
STATUS SETUP
Ethernet enable
Allows the user to
enable
or
disable
the Ethernet interface, to allow
communication between the analyzer and network. This should be
set disabled when the Ethernet is not in use to prevent unauthorized
network access.
To change this, press
ENTER
to enter Edit mode and then use the
UP
/
DOWN
arrow keys to toggle between
Enabled
or
Disabled
.
DHCP enable
Allows the user to
enable
or
disable
the DHCP (Dynamic Host Configu-
ration Protocol) client. This function should be disabled to ensure the
network uses a static address. The DHCP server must be on the same
Subnet as the analyzer.
IP Address
Defines the network of the analyzer. This is a required parameter. To
change this, press
ENTER
to enter Edit mode and change this address.
IP Subnet Mask
A TCP/IP number used to determine to which the TCP/IP Subnet de-
vice belongs. Devices in the same Subnet can be communicated with
locally without going through a router. This is a required parameter.
Ethernet Network Screen
The
Ethernet Network
screen (Figure 4-36) allows the user to enter Ether-
net settings. This is used to enable your analyzer system to communicate
over an Ethernet network for viewing analyzer and stream composition
data, viewing analyzer diagnostics, or backing up/restoring analyzer con-
figuration settings. For addresses, contact your network administrator.
When changing the
IP Address
,
IP Subnet Mask
, and
Gateway
IP ad-
dress, use the numeric keys to enter a new value. When a period ( . )
character is necessary, after every third digit of an address, the system will
only accept input from the decimal key ( ‘
.
’ ). Use the
LEFT
arrow key as a
back space and to delete numerical characters to the left of the cursor, or
the
RIGHT
arrow key to advance to the right of the cursor.
Содержание IPS-4-UV
Страница 15: ...Contents xv...
Страница 16: ...xvi IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 24: ...1 8 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 66: ...3 36 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 138: ...5 32 IPS 4 UV Analyzer GPDiv2 Figure 5 4 Over Temperature alarm Reset Switch S501 Relay board 100 1942...
Страница 144: ...6 6 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 180: ...A 2 IPS 4 UV Analyzer GPDiv2 Analyzer Light Path Schematic...
Страница 183: ...Appendix Drawings A 5 Detector Interface Board 100 1937...
Страница 184: ...A 6 IPS 4 UV Analyzer GPDiv2 Detector Board 100 1955...
Страница 185: ...Appendix Drawings A 7 Display Interface Board 100 1978...
Страница 186: ...A 8 IPS 4 UV Analyzer GPDiv2 Customer Connection Board 100 1971...
Страница 187: ...Appendix Drawings A 9 Relay Board 100 1942...
Страница 188: ...A 10 IPS 4 UV Analyzer GPDiv2 Analog Board 100 1922...
Страница 189: ...Appendix Drawings A 11 MCU Board 100 1914...
Страница 190: ...A 12 IPS 4 UV Analyzer GPDiv2 Xenon Lamp Board 100 2061...
Страница 191: ...Appendix Drawings A 13 Electronics Enclosure Wiring Diagram Sheet 1 of 3 WX 102877...
Страница 192: ...A 14 IPS 4 UV Analyzer GPDiv2 Electronics Enclosure Wiring Diagram Sheet 2 of 3 WX 102877...
Страница 193: ...Appendix Drawings A 15 Electronics Enclosure Wiring Diagram Sheet 3 of 3 WX 102877...
Страница 194: ...A 16 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 196: ...S 2 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 200: ...G 4 IPS 4 UV Analyzer GPDiv2 g 1035 17 32 Not used 1036 Not used 1037 Not used 1038 Not used 1039 Not used...