4-24 | IPS-4-UV Analyzer-GPDiv2
NONE
Select NONE if you do not wish to assign a relay to an output
channel.
Analyte 1–8
Defines the analyte (species) of interest for the relay output.
You can assign a different relay for each channel. To assign a relay
to a channel, select an analyte and press
ENTER
. Note that the
actual species name (e.g., SO
2
) will be displayed on the screen, not
“Analyte.”
Relay 1–8 Value
Defines the
Low
or
High Alarm Limit
threshold for each relay (user
input required). This parameter only applies if
Relay Function
is set to
Low Limit
or
High Limit
.
Relay 1–8 Delay
Defines the delay time to de-energize the relay after the analyte con-
centration goes outside the limit set with
Relay Value
.
For example, if
Relay Delay
is set to 5 seconds, the analyte concentra-
tion can exceed the High Limit for up to 5 seconds at a time without
de-energizing the relay. If the concentration stays high for more than
5 seconds, the relay will be de-energized. The range is 0–3600 seconds.
This parameter only applies if
Relay Function
is set to
Low Limit
or
High Limit
.
To change the settings for a relay:
1. Enter a password to allow changes to be made to the relay settings.
2. Select one of the
Relay
(
1–8
) sub-menus. Press
ENTER
to view the pa-
rameters for that relay.
3. Select the required parameter (
Relay
Function, Source, Value,
or
Delay
),
change the settings, and press
ENTER
.
Repeat this step for each parameter and each relay.
When entering a value for the
Relay
Value
and
Delay
parameters:
•
Use the
UP
/
DOWN
arrow key to increase/decrease the value in
increments of one (1.00). This works best for minor changes. The
relay value is displayed on the right side of the screen.
•
If a value exists, use the
LEFT
arrow key as a backspace to erase the
characters to the left of the cursor ( ‘
_
’ ) if applicable, then use the
numeric keys to enter the new value. This works best for drastic
changes.
Содержание IPS-4-UV
Страница 15: ...Contents xv...
Страница 16: ...xvi IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 24: ...1 8 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 66: ...3 36 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 138: ...5 32 IPS 4 UV Analyzer GPDiv2 Figure 5 4 Over Temperature alarm Reset Switch S501 Relay board 100 1942...
Страница 144: ...6 6 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 180: ...A 2 IPS 4 UV Analyzer GPDiv2 Analyzer Light Path Schematic...
Страница 183: ...Appendix Drawings A 5 Detector Interface Board 100 1937...
Страница 184: ...A 6 IPS 4 UV Analyzer GPDiv2 Detector Board 100 1955...
Страница 185: ...Appendix Drawings A 7 Display Interface Board 100 1978...
Страница 186: ...A 8 IPS 4 UV Analyzer GPDiv2 Customer Connection Board 100 1971...
Страница 187: ...Appendix Drawings A 9 Relay Board 100 1942...
Страница 188: ...A 10 IPS 4 UV Analyzer GPDiv2 Analog Board 100 1922...
Страница 189: ...Appendix Drawings A 11 MCU Board 100 1914...
Страница 190: ...A 12 IPS 4 UV Analyzer GPDiv2 Xenon Lamp Board 100 2061...
Страница 191: ...Appendix Drawings A 13 Electronics Enclosure Wiring Diagram Sheet 1 of 3 WX 102877...
Страница 192: ...A 14 IPS 4 UV Analyzer GPDiv2 Electronics Enclosure Wiring Diagram Sheet 2 of 3 WX 102877...
Страница 193: ...Appendix Drawings A 15 Electronics Enclosure Wiring Diagram Sheet 3 of 3 WX 102877...
Страница 194: ...A 16 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 196: ...S 2 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 200: ...G 4 IPS 4 UV Analyzer GPDiv2 g 1035 17 32 Not used 1036 Not used 1037 Not used 1038 Not used 1039 Not used...