Controller / User Interface | 4-21
Inject Zero fluid
This function injects Zero fluid into the sample system to flush the
Measuring Cell of light-absorbing compounds. During this operation,
the zero solenoid is open and “Zero Hold” is displayed in the
Status
area of the
HOME
screen. Note that this mode does not actually Zero
the analyzer.
Inject Span fluid
This function injects Span fluid into the sample system, which can
be used to observe the analyzer’s response to species of known
concentration(s). During this operation, the span solenoid is open and
“Span Hold” is displayed in the
Status
area of the
HOME
screen. Note
that this mode does not actually Span the analyzer.
Start measure
This function returns the analyzer to normal
Measure
mode from any
of the
Inject Zero fluid
,
Inject Span fluid
,
Zero Gas Flush
, or
Span
Gas Flush
functions. To return the analyzer to
Measure
mode, select
this function and press
ENTER
. The message
“Please wait .
..” appears
while the analyzer performs a Sample Flush before switching back to
Measure
mode.
Manually Starting a Zero Calibration
Refer to “Manually Zeroing the Analyzer Sample System” in Chapter 3 for
information about how to perform a manual zero on the analyzer.
Manually Starting a Span Calibration
This section applies only to systems with multi-species spanning capabil-
ity.
Pressing
CANCEL
at any time during this procedure will abort the
procedure and return the software to the
HOME
screen.
To introduce Span fluid:
1. Zero the analyzer, as described in Chapter 3 of this manual.
2. Ensure the Span fluid cylinder is connected to the Span Fluid Inlet on
the right side of the analyzer and is turned on.
For applications using multiple span fluids:
It is important to connect the appropriate Span fluid cylinder for the
corresponding fluid being spanned.
NOTE
Содержание IPS-4-UV
Страница 15: ...Contents xv...
Страница 16: ...xvi IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 24: ...1 8 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 66: ...3 36 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 138: ...5 32 IPS 4 UV Analyzer GPDiv2 Figure 5 4 Over Temperature alarm Reset Switch S501 Relay board 100 1942...
Страница 144: ...6 6 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 180: ...A 2 IPS 4 UV Analyzer GPDiv2 Analyzer Light Path Schematic...
Страница 183: ...Appendix Drawings A 5 Detector Interface Board 100 1937...
Страница 184: ...A 6 IPS 4 UV Analyzer GPDiv2 Detector Board 100 1955...
Страница 185: ...Appendix Drawings A 7 Display Interface Board 100 1978...
Страница 186: ...A 8 IPS 4 UV Analyzer GPDiv2 Customer Connection Board 100 1971...
Страница 187: ...Appendix Drawings A 9 Relay Board 100 1942...
Страница 188: ...A 10 IPS 4 UV Analyzer GPDiv2 Analog Board 100 1922...
Страница 189: ...Appendix Drawings A 11 MCU Board 100 1914...
Страница 190: ...A 12 IPS 4 UV Analyzer GPDiv2 Xenon Lamp Board 100 2061...
Страница 191: ...Appendix Drawings A 13 Electronics Enclosure Wiring Diagram Sheet 1 of 3 WX 102877...
Страница 192: ...A 14 IPS 4 UV Analyzer GPDiv2 Electronics Enclosure Wiring Diagram Sheet 2 of 3 WX 102877...
Страница 193: ...Appendix Drawings A 15 Electronics Enclosure Wiring Diagram Sheet 3 of 3 WX 102877...
Страница 194: ...A 16 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 196: ...S 2 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 200: ...G 4 IPS 4 UV Analyzer GPDiv2 g 1035 17 32 Not used 1036 Not used 1037 Not used 1038 Not used 1039 Not used...