PN 9000-246-VE, Rev D
Alarm Condition / Description / Corrective Action
Alarm Type
Cell Temp Hi
(User Interface)
Cell Temperature High
(5100P Application Software)
The temperature of the Measuring Cell exceeds the allowed operating range for the analyzer.
This condition exists when the analyzer is used in ambient temperatures that exceed the rated
range of 50 °C (122.0 °F) or if the temperature of the Sample gas supplied to the analyzer
exceeds the rated value.
Corrective Action:
• Protect the 5100P Analyzer from direct sunlight and/or operate it only when ambient
temperatures are within the defined range of -20 °C to +50 °C (-4 °F to 122.0 °F). If sample
temperatures are consistently near or greater than 50 °C (122.0 °F), efforts should be made
to cool the sample before it is analyzed in the 5100P.
Warning
Cell Temp Lo
(User Interface)
Cell Temperature Low
(5100P Application Software)
The temperature of the Measuring Cell drops below the allowed operating range for the ana-
lyzer. This condition exists when the instrument is used in ambient temperatures that exceed
the rated range of -20 °C (-4.0 °F), or if the temperature of the Sample gas supplied to the
analyzer is below the rated value.
Corrective Action:
• Protect the 5100P Analyzer from extreme cold and/or operate it only when ambient
temperatures are within the defined range of -20 °C to +50 °C (-4 °F to 122.0 °F). If sample
temperatures are consistently below -20 °C (-4.0 °F), efforts should be made to warm the
sample before it is analyzed in the 5100P.
Warning
System Fault
(User Interface and 5100P Application Software)
The 5100P has detected a failure in the
Detector
electronics.
Corrective Action:
Take appropriate safety precautions, and:
• Turn Off the power to the analyzer. Restart the 5100P. If the alarm message appears repeat-
edly, contact AMETEK Service and Support for assistance.
Fatal
Data Invalid
(User Interface and 5100P Application Software)
The 5100P is unable to lock on the absorption line. Values are invalid. This alarm can occur
during startup or if the analyzer is exposed to a sudden change in the ambient temperature.
The analyzer should re-establish and verify line-lock within a few moments and return to
normal operation.
Corrective Action:
• If the analyzer has been powered up for 3–5 minutes and local environmental and sample
temperatures are not changing significantly, the alarm should not persist. If the alarm mes-
sage appears repeatedly, contact AMETEK Service and Support for assistance.
Fatal
Maintenance and Troubleshooting
|
4-13
Содержание 5100P
Страница 1: ...5100P TDLAS Analyzer USER MANUAL PN 9000 246 VE Rev D 5000 016 PI REV A...
Страница 10: ...PN 9000 246 VE Rev D This page intentionally left blank x 5100P TDLAS Analyzer...
Страница 28: ...PN 9000 246 VE Rev D This page intentionally left blank 2 14 5100P TDLAS Analyzer...
Страница 64: ...PN 9000 246 VE Rev D This page intentionally left blank 3 36 5100P TDLAS Analyzer...
Страница 82: ...PN 9000 246 VE Rev D This page intentionally left blank 5 4 5100P TDLAS Analyzer...