PN 9000-246-VE, Rev D
Standard Pressure Sample Panel (Typical Operation)
The
Standard
5100P Sample Conditioning Panel (Figure 2-2.1) operates with a
raw Sample gas pressure of 1.03–17.2 BARG (15–250 PSIG).
Downstream of the Inlet Valve, a pressure regulator is used to reduce the
Sample gas pressure to the recommended operating pressure of the 5100P
Analyzer (0.4–0.7 BARG [6–10 PSIG]).
The Sample gas passes through the Filter (membrane separator) to remove
any entrained (trapped) liquids and particulate matter before the gas passes
through the Flow Meter.
Figure 2-2.1.
Sample Panel Flow Diagram,
Maximum 250 PSIG (Standard
version).
FI
SAMPLE PANEL
TOP VIEW
INLET VALVE
REGULATOR
GENIE FILTER
DRAIN VALVE
FLOWMETER
WITH VALVE
OUTLET FITTING
(TO INSTRUMENT)
GAUGE
MAX PANEL INLET PRESSURE: 250 PSI
INSTRUMENT INLET/PANEL OUTLET: 101 KPA +/- 2.5 KPA
FLOW RATE: 1-2 SLPM
REGULATOR
INLET VALVE
GAUGE
DRAIN
VALVE
GENIE
FILTER
FLOWMETER
WITH VALVE
DRAIN
SAMPLE PANEL
FLOW SCHEMATIC
5100P
ANALYZER
6–10 PSIG
250 PSIG