PN 9000-246-VE, Rev D
2. If required, connect the International Power Adaptor to the Battery Char-
ger’s power cord and use the appropriate line plug adapter (i.e., US, Euro-
pean, etc.) to plug the cord into a power outlet.
3. After the charger has been connected to the AC mains voltage, one of two
LEDs will turn on:
• The YELLOW LED indicates that the charger is in
Charging
mode.
• The GREEN LED indicates that the charger is connected to power.
• When the Battery Pack is fully charged, the GREEN LED will be On and
the YELLOW LED will be Off.
4. Charge the 5100P for a minimum of eight (8) hours, (or overnight, if pos-
sible) prior to operating the analyzer using its Battery Pack. This should
provide 8-plus hours of continuous analyzer use. If required, the 5100P can
remain connected to the charger for an indefinite period while charging.
Extending Battery Life
Because the 5100P Analyzer is designed to be operated from the internal bat-
tery pack, it is important to operate the analyzer in a manner consistent with
obtaining maximum usable battery life between charges. If used properly, a
fully charged battery will provide a minimum of eight (8) hours of operation.
Consider the following measures to maximize the operational time on a fully
charged battery:
• Turn Off the power when the analyzer is not being used to measure
Sample gas.
• The backlight for the display will time out and turn Off automatically, ac-
cording to the
Backlight Timer
setting. Set the
Backlight Timer
to the
shortest practical duration to avoid draining the battery. See “Changing
the Backlight Time” in Chapter 3 for information on how to set this timer.
• Temperature extremes require extra power to run the TDL spectrometer
circuitry.
The battery pack will last longer when the analyzer is operated at or
near room temperature.
• Do not store the 5100P for longer than three (3) months (90 days) without
using the analyzer, or performing a full charging cycle. Always fully charge
the battery pack (minimum of 8 hours) prior to using the analyzer.
2-4 | 5100P TDLAS Analyzer
Содержание 5100P
Страница 1: ...5100P TDLAS Analyzer USER MANUAL PN 9000 246 VE Rev D 5000 016 PI REV A...
Страница 10: ...PN 9000 246 VE Rev D This page intentionally left blank x 5100P TDLAS Analyzer...
Страница 28: ...PN 9000 246 VE Rev D This page intentionally left blank 2 14 5100P TDLAS Analyzer...
Страница 64: ...PN 9000 246 VE Rev D This page intentionally left blank 3 36 5100P TDLAS Analyzer...
Страница 82: ...PN 9000 246 VE Rev D This page intentionally left blank 5 4 5100P TDLAS Analyzer...