PN 9000-246-VE, Rev D
Connecting and Preparing the 5100P Analyzer
For normal operation, position the Sample Conditioning Panel vertically, so
that the Inlet Valve is located at the top of the panel. The analyzer should be
standing vertically on the supporting structure – with the User Interface dis-
play facing upward – so that the flow of Sample gas is correctly metered.
Hazardous Locations
• Do not connect the 5100P Analyzer to a computer (via the USB
port).
• To meet Hazardous Location and Ingress Protection ratings,
ensure the protective caps for the Battery Charger and USB Data
Connection ports (on the 5100P Rear Panel) are securely installed
on their respective ports.
• When working with the 5100P Analyzer in ATEX-IECEx hazardous
locations, use only the built-in Battery Pack to power the analyzer.
Do not connect and use the Battery Charger to supply power to
the analyzer.
• Ensure the area is safe for the operation of this equipment.
• Ensure all personnel performing work with the analyzer and
Sample gas stream are familiar with all applicable safe handling
procedures.
• Ensure all required Personal Protective Equipment (PPE) is being
appropriately used.
• Ensure the drain connection of the Sample Panel and the outlet
of the analyzer are connected to a safe venting point that is at
atmospheric pressure.
Before connecting Sample gas and operating the 5100P Analyzer and
its Sample Conditioning Panel, read the entire section “Introduction to
the User Interface” in Chapter 3 to learn how to navigate through the
screens to work on the analyzer.
Connecting and Starting Up the 5100P
|
2-9
Содержание 5100P
Страница 1: ...5100P TDLAS Analyzer USER MANUAL PN 9000 246 VE Rev D 5000 016 PI REV A...
Страница 10: ...PN 9000 246 VE Rev D This page intentionally left blank x 5100P TDLAS Analyzer...
Страница 28: ...PN 9000 246 VE Rev D This page intentionally left blank 2 14 5100P TDLAS Analyzer...
Страница 64: ...PN 9000 246 VE Rev D This page intentionally left blank 3 36 5100P TDLAS Analyzer...
Страница 82: ...PN 9000 246 VE Rev D This page intentionally left blank 5 4 5100P TDLAS Analyzer...