PN 9000-246-VE, Rev D
Refresh
button
If changes have been made to any of the values or settings under
Alarms
,
Background Gas
, or
Parameters
but the
Set
button has not been
pressed, press the
Refresh
button to return the settings to their original
values or settings.
The
Calibration
group allows you to configure the Span Concentration and
initiate a Span or Zero Calibration. For details about calibrating the system,
refer to "Span or Zero Calibrating the 5100P Analyzer" in this chapter.
The
Backup/Restore
group allows you to save (back up) the current analyzer
configuration settings for future use or to return modified settings to their
original settings. You can also restore the original analyzer settings from a
file that was previously backed up. For details about backing up or restor-
ing analyzer settings, refer to "Backing Up and Restoring 5100P Analyzer
Settings" in this chapter.
3-32 | 5100P TDLAS Analyzer