PN 9000-246-VE, Rev D
Connecting Sample Gas to the 5100P Sample Panel
Before connecting the Sample gas to the Sample Conditioning Panel and ana-
lyzer, verify the operating pressure of the Sample gas line does not exceed the
maximum rating of the Sample Panel. Check the label on the analyzer’s Sample
Panel to verify the maximum pressure rating for your system.
To connect the Sample gas line and operate the Sample Panel:
1. Before beginning, close the Sample Panel’s Inlet Valve to isolate the 5100P
Sample Conditioning Panel from the sample line (see Figures 2-2.1 and
2-2.2 for location of Inlet Valve). The Inlet Valve is closed when it is in the
vertical position (when the 5100P is also in the vertical position).
Systems with the
Standard
(1.03–17.2 BARG/15–250 PSIG) Sample
Conditioning Panel only:
Fully close the Regulator on the AMETEK Sample Conditioning Panel.
(Fully clockwise = Closed.)
2. Connect the sample line to the 1/8-inch compression fitting on the Sample
Panel’s Inlet Valve.
3. Ensure there are no leaks at this connection. Use Snoop® or another suit-
able leak detection agent to check the connections.
4. Open the main process Isolation Valve (by customer/end user) on the pro-
cess pipe to start the flow of Sample gas to the Sample Panel.
5. Open the Drain Valve 1/8-turn by turning the valve 1/8-turn counter-clock-
wise from the fully closed position.
6. Turn the Inlet Valve on the Sample Panel to the Open position.
7. Using the Regulator, adjust the sample pressure to between 0.4–0.7 BARG
(6–10 PSIG).
8. Adjust the Flow Meter Valve to set the Sample gas flow to between 1–2
LPM (see Figures 2-2.1 and 2-2.2 for location of Flow Meter with Valve).
9. After the flow rate of the Sample gas has stabilized to 1–2 LPM through the
5100P, the analyzer is ready to measure the analyte concentration in the
gas stream. See “Start-Up and Verification” in this chapter and Chapter 3 for
information about operating the analyzer.
2-10 | 5100P TDLAS Analyzer
Содержание 5100P
Страница 1: ...5100P TDLAS Analyzer USER MANUAL PN 9000 246 VE Rev D 5000 016 PI REV A...
Страница 10: ...PN 9000 246 VE Rev D This page intentionally left blank x 5100P TDLAS Analyzer...
Страница 28: ...PN 9000 246 VE Rev D This page intentionally left blank 2 14 5100P TDLAS Analyzer...
Страница 64: ...PN 9000 246 VE Rev D This page intentionally left blank 3 36 5100P TDLAS Analyzer...
Страница 82: ...PN 9000 246 VE Rev D This page intentionally left blank 5 4 5100P TDLAS Analyzer...