PN 9000-246-VE, Rev D
To install the o-ring (new or original), lay it on the Membrane and center
it over its o-ring groove. Using your thumb or finger, gently push one side
into its o-ring groove, then repeat on the opposite side. Continue to press
down the o-ring into its groove, alternating sides until the o-ring is seated
and the Filter Membrane is secure in the Filter Cap.
Finally, turn the Filter Cap over and press it downward on a clean, flat sur-
face to evenly seat the o-ring in its groove.
7. Replace the Filter Cap on the Filter Membrane and hand-tighten it.
Do not
use a wrench to tighten it.
Systems Without an AMETEK Sample Panel
5100P Analyzers that will be used in ATEX/IECEx Zone 2 areas must
have an AMETEK-supplied Sample Conditioning Panel.
If you are operating the 5100P Analyzer without an AMETEK-supplied Sample
Conditioning Panel, precautions must be taken to remove liquid and solid par-
ticulates from the Sample gas, to ensure analyzer performance and accuracy.
Failure to provide a properly conditioned Sample gas can result in
physical damage to the 5100P and may also result in incorrect analyte
concentration measurements. Refer to “No Sample Panel” in Chapter 2
for more information.
This can be accomplished by using appropriate filters (supplied by Customer).
AMETEK recommends using a 2-micron particulate filter and a filter membrane
capable of removing any liquids, including droplets, or other contaminants in
the sample stream. For new installations, the filters must be checked frequent-
ly (e.g., weekly) to determine how long it typically takes for the filter to become
dirty. If the filter remains clean for extended periods of time, the filter can be
checked less frequently, if preferred.
4-10 | 5100P TDLAS Analyzer
Содержание 5100P
Страница 1: ...5100P TDLAS Analyzer USER MANUAL PN 9000 246 VE Rev D 5000 016 PI REV A...
Страница 10: ...PN 9000 246 VE Rev D This page intentionally left blank x 5100P TDLAS Analyzer...
Страница 28: ...PN 9000 246 VE Rev D This page intentionally left blank 2 14 5100P TDLAS Analyzer...
Страница 64: ...PN 9000 246 VE Rev D This page intentionally left blank 3 36 5100P TDLAS Analyzer...
Страница 82: ...PN 9000 246 VE Rev D This page intentionally left blank 5 4 5100P TDLAS Analyzer...