PN 9000-246-VE, Rev D
Span or Zero Calibrating the 5100P Analyzer
The 5100P Analyzer calibration functions are initiated from the
Setup
screen.
Span Calibrating the 5100P Analyzer
To perform a Span Calibration on the 5100P Analyzer:
1. From the
Setup
screen, enter the
Span Concentration
value (from the
Span gas bottle).
2. Click the
Span
button. The Span Calibration will begin and the “
Calibration
is in progress
" status bar is displayed below the
Calibration
group. Note
that the
Span
and
Zero
buttons change to
Accept span
and
Cancel span
.
Wait for the
Concentration
reading to stabilize and then click the
Accept
span
button. The message “
Click YES to accept this span calibration. NO
to cancel.
” will appear. To accept the Span Calibration, click
Yes
. To cancel
the Span Calibration, click
No
; the calibration will stop and the analyzer
will not accept the Span Calibration.
Figure 3-21.1.
Span or Zero Calibrating the
5100P, Setup screen.
Figure 3-21.2.
Span Calibrating the 5100P,
Setup screen.
User Interface/5100P Application Software
|
3-33