Configuring the 5100P Parameters From the Setup Screen
The
Setup
screen displays system version information and allows you to
configure several
Parameters
,
Alarms
,
Background Gas
, and analyte mea-
surement
Units
. When you select
Setup
, the software will read the analyzer
parameter values and update the information on this screen.
The
System
group displays firmware versions of the analyzer’s
MCU
and
DSP
boards, and the version of the
Windows Software
(5100P Application
Software).
The
Alarms
group allows you to configure the upper and lower output limits
for the Concentration alarms. These values must correspond with the
Units
Concentration Value
selected.
Output High Limit
This is the value at which a
Concentration High
alarm will be trig-
gered if the concentration of the analyte being measured exceeds this
value. To change the
High Limit
value, click in the numerical text field
and enter a new value. Click
Set
to save the change.
Output Low Limit
This is the value at which a
Concentration Low
alarm will be triggered
if the concentration of the analyte being measured falls below this
value. To change the
Low Limit
value, click in the numerical text field
and enter a new value. Click
Set
to save the change.
Background Gas
Depending on the analyte being measured the background gas option(s)
will differ. To change the background gas, choose an option from the
drop-down list. Click
Set
to save the change. Refer to the accompanying
analyte-specific
Manual Supplement
shipped with this manual for addi-
tional information.
Figure 3-20.
Setup screen (Moisture analyzer
example screen shown).
3-30 | 5100P TDLAS Analyzer
Содержание 5100P
Страница 1: ...5100P TDLAS Analyzer USER MANUAL PN 9000 246 VE Rev D 5000 016 PI REV A...
Страница 10: ...PN 9000 246 VE Rev D This page intentionally left blank x 5100P TDLAS Analyzer...
Страница 28: ...PN 9000 246 VE Rev D This page intentionally left blank 2 14 5100P TDLAS Analyzer...
Страница 64: ...PN 9000 246 VE Rev D This page intentionally left blank 3 36 5100P TDLAS Analyzer...
Страница 82: ...PN 9000 246 VE Rev D This page intentionally left blank 5 4 5100P TDLAS Analyzer...