PN 9000-246-VE, Rev D
Alarm Condition / Description / Corrective Action
Alarm Type
Battery Lo
(User Interface)
Battery Low
(5100P Application Software)
The 5100P’s internal battery needs to be charged. There is approximately 5–10 minutes of
analyzer run-time before it will power itself down completely. Typically, a minimum of eight
(8) hours of charging (or overnight, if possible) should provide 8-plus hours of continuous
analyzer use.
Corrective Action:
Take appropriate safety precautions, and:
• Using the supplied Battery Charger, charge the analyzer in an approved and safe location
prior to use. If the analyzer Low Battery alarm persists after charging, contact AMETEK Ser-
vice and Support for assistance.
Warning
Clipping Spectra
(5100P Application Software)
The 5100P’s
Sample
and/or
Reference
Detector voltage exceeds expected range.
Corrective Action:
Contact AMETEK Service and Support for assistance.
Warning
Low Signal (Power)
(5100P Application Software)
The 5100P the signal from the Optical Cell is too low.
Corrective Action:
Contact AMETEK Service and Support for assistance.
Warning
Calibration Invalid
(5100P Application Software)
The factory calibration has become corrupted and the results are no longer valid.
Corrective Action:
Contact AMETEK Service and Support for assistance.
Fatal
4-14 | 5100P TDLAS Analyzer
Содержание 5100P
Страница 1: ...5100P TDLAS Analyzer USER MANUAL PN 9000 246 VE Rev D 5000 016 PI REV A...
Страница 10: ...PN 9000 246 VE Rev D This page intentionally left blank x 5100P TDLAS Analyzer...
Страница 28: ...PN 9000 246 VE Rev D This page intentionally left blank 2 14 5100P TDLAS Analyzer...
Страница 64: ...PN 9000 246 VE Rev D This page intentionally left blank 3 36 5100P TDLAS Analyzer...
Страница 82: ...PN 9000 246 VE Rev D This page intentionally left blank 5 4 5100P TDLAS Analyzer...