PN 9000-246-VE, Rev D
Figure 3-5.1.
MENU screen, ALARM
sub-menu selected.
MENU
( ALARM )
CONFIG
Figure 3-5.2.
ALARM CONCENTRATION
HI/LO LIMIT sub-menu.
CONCENTRATION
( LO LIMIT )
HI LIMIT
Working From the Menu Screen
Working from the 5100P Analyzer User Interface
MENU
screen is limited to two
options, one for editing alarm concentration setpoints, and one for configur-
ing analyzer and application parameters. To select a menu item, use the
Up
or
Down
arrow key to toggle between the two options. The parentheses indicate
the selected item.
Editing Alarm Concentration Low/High Setpoints
The 5100P can be changed to enable concentration alarm messages. These
messages will be displayed on the User Interface screen and recorded in the
Data Log.
To set alarm limits:
1. From the
HOME
screen, press the
Enter
key to view the
MENU
screen. The
menu items
ALARM
and
CONFIG
are displayed. The default selection is
ALARM
.
2. Press
Enter
to view the
ALARM CONCENTRATION
screen, then press
Enter
again to view the
HI/LO LIMIT
sub-menu.
3-6 | 5100P TDLAS Analyzer