PN 9000-246-VE, Rev D
No Sample Panel
5100P Analyzers that will be used in ATEX/IECEx Zone 2 areas must
have an AMETEK-supplied Sample Conditioning Panel.
When operating the 5100P Analyzer without an AMETEK-supplied Sample
Conditioning Panel, precautions must be taken to ensure the accuracy of
performance and reduce possible maintenance issues.
Ensure the following conditions are met:
• Sample flow rate needs to be controlled between 1–2 LPM.
• The Sample Inlet pressure must be 0.4–0.7 BARG (6–10 PSIG).
• Liquid and solid particulates must be removed from the Sample gas by us-
ing appropriate filters (supplied by Customer). AMETEK recommends using
a 2-micron particulate filter and a filter membrane capable of removing
any liquids, including droplets.
Fail
ure to provide a properly conditioned Sample gas can result
in physical damage to the 5100P and may also result in incorrect
analyte concentration measurements.
2-8 | 5100P TDLAS Analyzer