Agilent 1260 Infinity VWD User Manual
4
Using the Detector
Special Settings of the Detector
Scanning with the VWD
1
Set up a run.
2
Start a run.
3
While running on the baseline, select from the menu
Instrument
>
More VWD
>
Blank Scan
.
A background scan is stored in the memory.
4
When the peak of interest enters the flow cell, stop the flow (set flow
rate to zero or open the purge valve) and wait a few moments to
stabilize the concentration.
5
Select from the menu
Instrument > More VWD > Sample Scan
.
A sample scan is taken in the range defined under
on page 70 and the Online Spectra window (see
on page 67) displays the result (Sample Scan minus Blank
Scan).
•
Step 1:
Blank Scan
: scan of the background (solvent)
is stored in the memory.
•
Step 2:
Sample Scan
: scan of the peak of interest is
taken while the peak stays in the flow cell (stop-flow
condition).
•
Online Spectrum
:
Sample Scan
minus
Blank Scan
.
N O T E
Turning off the pump would stop the run and no access to the sample scan is possible.
Содержание G1314B
Страница 1: ...Agilent Technologies Agilent 1260 Infinity Variable Wavelength Detector User Manual...
Страница 30: ...30 Agilent 1260 Infinity VWD User Manual 2 Site Requirements and Specifications Performance Specifications...
Страница 102: ...102 Agilent 1260 Infinity VWD User Manual 7 Error Information Detector Error Messages...
Страница 164: ...164 Agilent 1260 Infinity VWD User Manual 11 Identifying Cables External Contact Cable...
Страница 190: ...190 Agilent 1260 Infinity VWD User Manual 12 Hardware Information Early Maintenance Feedback EMF...
Страница 207: ......