Agilent 1260 Infinity VWD User Manual
4
Using the Detector
Special Settings of the Detector
Special Setpoints
1
To change the offset and the attenuation select
VWD Signal > More >
Special Setpoints
.
Figure 28
Spectra Window
•
Margin for negative Absorbance:
Use this field to modify the detector’s
signal handling to increase the margin
for negative absorbance. Use this
option if, for example, your solvent
gradient produces a decreasing
baseline absorbance, and for GPC
analyses.
Limits: 100 – 4000 mAU.
•
Signal Polarity:
can be switched to
negative (if required).
•
Enable analysis when lamp is off:
if
the VWD is not used in a dual detector
setup (lamp off), the not-ready
condition is not stopping the analysis.
•
Scan Range / Step:
Used for
stop-flow scanning, see
N O T E
Margin for negative Absorbance
: The higher the value the greater the baseline noise. Set
this value only if you expect negative absorbance greater than -100 mAU.