130
Agilent 1260 Infinity VWD User Manual
9
Maintenance
Using the Cuvette Holder
Using the Cuvette Holder
This cuvette holder can be placed instead of a flow cell in the variable
wavelength detector. Standard cuvettes with standards in it, for example,
National Institute of Standards & Technology (NIST) holmium oxide
solution standard, can be fixed in it.
This can be used for wavelength verifications.
When
If your own standard should be used to checkout the instrument.
Parts required
#
p/n
Description
1
G1314-60200
Cuvette Holder
1
Cuvette with the “standard”, e.g. NIST certified holmium oxide sample
Preparations
•
Remove the normal flow cell.
•
Have cuvette with standard available.