Agilent 1260 Infinity VWD User Manual
127
Maintenance
9
Exchanging a Flow Cell
3
Unscrew the thumb screws parallel and remove the flow
cell.
N O T E
If you want to maintain flow cell parts, see
on page 128or the information
provided with your flow cell.
4
Press the flow cell completely into the slot and tighten
the cell screws (both parallel) until the mechanical stop.
Reconnect the inlet and outlet capillaries to the flow cell.
5
Replace the front cover.
Next Steps:
6
To check for leaks, establish a flow and observe the flow
cell (outside of the cell compartment) and all capillary
connections.
7
Insert the flow cell.
8
Perform
“Wavelength Verification-Calibration”
page 108 to check the correct positioning of the flow cell.
9
Replace the front cover.