Agilent 1260 Infinity VWD User Manual
105
Test Functions
8
Intensity Test
Instensity Test with Agilent Lab Advisor
Figure 33
Intensity Test with Agilent Lab Advisor
Intensity Test Failed
Probable cause
Suggested actions
1
Empty flow cell
Ensure the flow cell is filled with water.
2
Flow cell windows dirty
Repeat the test with the flow cell removed. If
the test passes, exchange the flow cell
windows.
3
Optics defect
Please contact your Agilent service
representative.
4
Defective lamp or optics.
Exchange the lamp.