132
Agilent 1260 Infinity VWD User Manual
9
Maintenance
Correcting Leaks
Correcting Leaks
1
Remove the front cover.
2
Use tissue to dry the leak sensor area.
3
Observe the capillary connections and the flow cell area for leaks and
correct, if required.
4
Replace the front cover.
Figure 42
Drying the Leak Sensor
When
If a leakage has occurred in the flow cell area or at the capillary connections.
Tools required
Description
Tissue
Wrench, 1/4 inch
for capillary connections