Agilent 1260 Infinity VWD User Manual
63
Using the Detector
4
Setting up an Analysis
The Online Plot (
on page 63) shows both, the pump pressure
and the detector absorbance signals. Pressing
Adjust
the signals can be
reset to the offset value and
Balance
would do a balance on the
detector.
Figure 21
Online Plot Window
13
If both baselines are stable, set the Y- range for the detector signal to
100 mAU.
EjbeegZhhjgZh^\cVa
KL9VWhdgWVcXZh^\cVa
N O T E
If you start with a new UV-lamp for the first time, the lamp may show initial drift for some
time (burn-in effect).