Agilent 1260 Infinity VWD User Manual
133
Maintenance
9
Replacing Leak Handling System Parts
Replacing Leak Handling System Parts
1
Remove the front cover to have access to the leak handling system.
2
Pull the leak funnel out of the leak funnel holder.
3
Pull the leak funnel with the tubing out of its location.
4
Replace the leak funnel and/or the tubing.
5
Insert the leak funnel with the tubing in its position.
6
Insert the leak funnel into the leak funnel holder.
7
Replace the front cover.
Figure 43
Replacing Waste Handling System Parts
When
If the parts are corroded or broken.
Tools required
None
Parts required
#
p/n
Description
1
5041-8389
Leak funnel holder
1
5041-8388
Leak funnel
1
5062-2463
Corrugated tubing, PP, 6.5 mm id, 5 m
AZV`[jccZa]daYZg
AZV`[jccZa
AZV`ijW^c\
AZV`eaVc
LVhiZdjiaZi
AZV`hZchdgVhhZbWan