Agilent 1260 Infinity VWD User Manual
69
Using the Detector
4
Special Settings of the Detector
Analog Output Settings
1
To change the Output Range of the analog outputs select
VWD Control
.
2
To change the offset and the attenuation select
VWD Signal
>
More
.
3
Change the values if required.
Figure 26
Analog Output Settings
•
Analog Output Range
: can be set to either 100 mV
or 1 V full scale.
•
Zero Offset
:can be set to either 100 mV or 1 V full
scale.
•
Attenuation Limits
: 0.98 to 4000 mAU at discrete
values for either 100 mV or 1 V full scale.
Figure 27
Analog Output Settings