Agilent 1260 Infinity VWD User Manual
12
Hardware Information
Setting the 8-bit Configuration Switch (without On-board) LAN
Forced Cold Start
A forced cold start can be used to bring the module into a defined mode
with default parameter settings.
If you use the following switch settings and power the instrument up
again, a forced cold start has been completed.
C A U T I O N
Loss of data
Forced cold start erases all methods and data stored in the non-volatile memory.
Exceptions are calibration settings, diagnosis and repair log books which will not be
erased.
➔
Save your methods and data before executing a forced cold start.
Table 26
Forced Cold Start Settings (without on-board LAN)
Mode Select
SW1
SW2
SW3
SW4
SW5
SW6
SW7
SW8
TEST/BOOT
1
1
0
0
0
0
0
1