116
Agilent 1260 Infinity VWD User Manual
8
Test Functions
Holmium Oxide Test
Holmium Oxide Test Failed
Probable cause
Suggested actions
1
Detector not calibrated.
Recalibrate the detector.
2
Dirty or defective flow cell.
Repeat the test with the flow cell removed. If
the test is OK, exchange the flow cell
components.
3
Dirty or defective holmium oxide filter.
Run the holmium oxide filter test. If the test
fails, contact your Agilent service
representative.
4
Optical misalignment.
Please contact your Agilent service
representative.