Agilent 1260 Infinity VWD User Manual
99
Error Information
7
Detector Error Messages
Wavelength test failed
Error ID: 7890
The automatic wavelength check after lamp ignition has failed. When the
lamp is switched on, the detector waits 1 min to warm- up the lamp. Then
a check of the deuterium emission line (656 nm) via the reference diode is
performed. If the emission line is more than 3 nm away from 656 nm, the
error message is generated.
Probable cause
Suggested actions
1
Motor is not connected.
Please contact your Agilent service
representative.
2
Defective motor.
Please contact your Agilent service
representative.
3
Defective/missing grating or filter.
Please contact your Agilent service
representative.
4
Cable/connector defective.
Please contact your Agilent service
representative.
Probable cause
Suggested actions
1
Calibration incorrect.
Recalibrate the detector.
Содержание G1314B
Страница 1: ...Agilent Technologies Agilent 1260 Infinity Variable Wavelength Detector User Manual...
Страница 30: ...30 Agilent 1260 Infinity VWD User Manual 2 Site Requirements and Specifications Performance Specifications...
Страница 102: ...102 Agilent 1260 Infinity VWD User Manual 7 Error Information Detector Error Messages...
Страница 164: ...164 Agilent 1260 Infinity VWD User Manual 11 Identifying Cables External Contact Cable...
Страница 190: ...190 Agilent 1260 Infinity VWD User Manual 12 Hardware Information Early Maintenance Feedback EMF...
Страница 207: ......