Replacing the Module’s Firmware
Replacing the Module’s Firmware
To upgrade/downgrade the module’s firmware carry out the following
steps:
1
Download the required module firmware, the latest LAN/RS- 232 FW
Update Tool and the documentation from the Agilent web.
•
http://www.chem.agilent.com/_layouts/agilent/downloadFirmware.aspx?whid=69761
2
For loading the firmware into the module follow the instructions in the
documentation.
When
The installation of newer firmware might be necessary
•
if a newer version solves problems of older versions or
•
to keep all systems on the same (validated) revision.
The installation of older firmware might be necessary
•
to keep all systems on the same (validated) revision or
•
if a new module with newer firmware is added to a system or
•
if third party control software requires a special version.
Tools required
Description
LAN/RS-232 Firmware Update Tool
OR
Agilent Lab Advisor software
OR
Instant Pilot G4208A
(only if supported by module)
Parts required
#
Description
1
Firmware, tools and documentation from Agilent web site
Preparations
Read update documentation provided with the Firmware Update Tool.
Содержание G1314B
Страница 1: ...Agilent Technologies Agilent 1260 Infinity Variable Wavelength Detector User Manual...
Страница 30: ...30 Agilent 1260 Infinity VWD User Manual 2 Site Requirements and Specifications Performance Specifications...
Страница 102: ...102 Agilent 1260 Infinity VWD User Manual 7 Error Information Detector Error Messages...
Страница 164: ...164 Agilent 1260 Infinity VWD User Manual 11 Identifying Cables External Contact Cable...
Страница 190: ...190 Agilent 1260 Infinity VWD User Manual 12 Hardware Information Early Maintenance Feedback EMF...
Страница 207: ......