Maintenance
9
Exchanging a Lamp
Exchanging a Lamp
When
If noise or drift exceeds application limits or lamp does not ignite
Tools required
Description
Screwdriver, Pozidriv #1 PT3
Parts required
p/n
Description
G1314-60100
Deuterium lamp
Preparations
Turn the lamp OFF.
WA R N I N G
Injury by touching hot lamp
If the detector has been in use, the lamp may be hot.
➔
If so, wait for lamp to cool down.
N O T E
If you want to use the Agilent DAD lamp instead of the VWD lamp, you have to change the
lamp settings in the VWD Configuration to lamp type 2140-0590. This ensures that the DAD
lamp’s filament heating is operated like in the DAD.
N O T E
The specification are based on Deuterium lamp (G1314-60100) and may be not achieved
when other lamp types or aged lamps are used.