Agilent 1260 Infinity VWD User Manual
43
Installing the Detector
3
Installing the Detector
Installing the Detector
For other cables see below and section
Instant Pilot (G4208A) is optional.
1
Install the LAN interface board in the detector (if required), see
“Replacing the Interface Board”
2
Place the detector in the stack or on the bench in a horizontal position.
Parts required
#
p/n
Description
1
Detector
1
Power cord
1
LAN cable (cross-over or twisted pair network cable)
1
Agilent ChemStation or other control software
1
G4208A
Instant Pilot
Preparations
Other LC modules must have appropriate firmware installed to work with the detector.
•
Locate bench space.
•
Provide power connections.
•
Unpack the detector.
N O T E
Before adding the detector into an existing system assure that the existing modules have
been updated to firmware revision that is supported by the control software.