Agilent 1260 Infinity VWD User Manual
9
Maintenance
Exchanging a Flow Cell
Exchanging a Flow Cell
When
If an application needs a different type of flow cell or the flow cell needs repair.
Tools required
Description
Wrench, 1/4 inch
for capillary connections
Parts required
#
p/n
Description
1
G1314-60186
Standard flow cell 10 mm, 14 µL, 40 bar
(with RFID tag)
OR
1
G1314-60187
Micro flow cell 3 mm, 2 µL, 120 bar
(with RFID tag)
OR
1
G1314-60183
Semi-micro flow cell 6 mm, 5 µL
(with RFID tag)
OR
1
G1314-60182
High pressure flow cell 10 mm, 14 µL, 400 bar
(with RFID tag)
Preparations
Turn the lamp OFF.
1
Press the release buttons and remove the front cover to
have access to the lamp area.
2
Disconnect the inlet and outlet capillaries.