Hardware Information
12
Setting the 8-bit Configuration Switch (without On-board) LAN
Communication Settings for RS-232C
The communication protocol used in the column compartment supports
only hardware handshake (CTS/RTR).
Switches 1 in down and 2 in up position define that the RS- 232C
parameters will be changed. Once the change has been completed, the
column instrument must be powered up again in order to store the values
in the non- volatile memory.
Use the following tables for selecting the setting which you want to use for
RS- 232C communication. The number 0 means that the switch is down
and 1 means that the switch is up.
N O T E
The following tables represent the configuration switch settings for the modules without
on-board LAN only.
Table 20
8-bit Configuration Switch (without on-board LAN)
Mode Select
1
2
3
4
5
6
7
8
RS-232C
0
1
Baudrate
Data
Bits
Parity
Reserved
1
0
Reserved
TEST/BOOT
1
1
RSVD
SYS
RSVD
RSVD
FC
N O T E
The LAN settings are done on the LAN Interface Card G1369B/C. Refer to the
documentation provided with the card.
Table 21
Communication Settings for RS-232C Communication (without on-board LAN)
Mode
Select
1
2
3
4
5
6
7
8
RS-232C
0
1
Baudrate
Data Bits
Parity
Содержание G1314B
Страница 1: ...Agilent Technologies Agilent 1260 Infinity Variable Wavelength Detector User Manual...
Страница 30: ...30 Agilent 1260 Infinity VWD User Manual 2 Site Requirements and Specifications Performance Specifications...
Страница 102: ...102 Agilent 1260 Infinity VWD User Manual 7 Error Information Detector Error Messages...
Страница 164: ...164 Agilent 1260 Infinity VWD User Manual 11 Identifying Cables External Contact Cable...
Страница 190: ...190 Agilent 1260 Infinity VWD User Manual 12 Hardware Information Early Maintenance Feedback EMF...
Страница 207: ......