Agilent 1260 Infinity VWD User Manual
131
Maintenance
9
Using the Cuvette Holder
1
Locate the cuvette holder on the desk.
2
Unscrew the bracket.
3
Insert the cuvette with the sample into the holder. The
clear side of the cuvette must be visible.
4
Replace the bracket and fix the cuvette.
Next Steps:
5
Install the cuvette holder in the instrument.
6
Perform your Wavelength Verification/Calibration (see
“Wavelength Verification-Calibration”
on page 108) to check the
correct position of the cuvette holder.