Agilent 1260 Infinity VWD User Manual
125
Maintenance
9
Exchanging a Lamp
5
Replace the front cover.
Next Steps:
6
Reset the lamp counter as described in the User Interface
documentation.
7
Turn the lamp ON.
8
Give the lamp more than 10 min to warm-up.
9
Perform
“Wavelength Verification-Calibration”
page 108 to check the correct positioning of the lamp.