Agilent 1260 Infinity VWD User Manual
189
Hardware Information
12
Early Maintenance Feedback (EMF)
Setting the EMF Limits
The setting of the EMF limits must be optimized over one or two
maintenance cycles. Initially, no EMF limit should be set. When instrument
performance indicates maintenance is necessary, take note of the values
displayed by lamp counters. Enter these values (or values slightly less
than the displayed values) as EMF limits, and then reset the EMF counters
to zero. The next time the EMF counters exceed the new EMF limits, the
EMF flag will be displayed, providing a reminder that maintenance needs
to be scheduled.
N O T E
This function is only available via Agilent Lab Advisor or Instant Pilot.