R3765/67G Series Network Analyzer Operation Manual
7.12 Limit Function
7-135
Edit limit menu (2 of 2)
{LIMIT LINE ON/OFF}
: Selects ON or OFF of the limit line display.
When the limit line is set and this is in the state of ON,
the limit line is displayed to compare the measurement
data on the scale.
The displays of the limit line are different depending on
DISPLAY FORMAT and LIMIT TYPE of the segment.
In the format of rectangular coordinates, and marks
or lines (straight lines or horizontal lines) are put
between the break points of each segment.
In polar coordinate, circle or straight line showing the
angle is drawn.
{LIMIT TEST ON/OFF}
: Selects ON/OFF of the limit test.
Under the limit line ON, the limit values and the data are
compared at each measurement point.
The limit test is performed when the data was updated
during sweeping or after swept, or when the limit test
was set to ON for the first time.
{BEEP [ ]}
: Calls the beep mode to set the beep for the limit test.
(See step
.)
{MAG DATA LIN/LOG}
: The limit test of Smith chart and polar display is judged
with MAG and PHASE.
Select whether this judgement is performed with LIN
(linear scale) or LOG (logarithmic scale). (Default setting
is LOG.)
This soft menu is effective only when the format (refer to
sub-section 7.4.2.) is of Smith chart or polar display.
{LIMIT MODE MENU}
: Calls the limit mode menu to control the limit test mode
and selects the limit type of polar coordinates format.
(See step
.)
{LIMIT LINE OFFSETS}
: Calls the offset limit menu to adjust the stimulus value
and the response value of the limit. (See step
.)
{CLEAR LIST}
: Calls the clear limit menu to clear all the segments in the
limit table. (See step
.)
7
4
5
11
9