R3765/67G Series Network Analyzer Operation Manual
3.5 Measurement Samples of Simple Transmission Characteristics
3-17
3.5.5
Group delay measurement
Setup (refer to sub-section 3.5.1) and preset (refer to sub-section 4.4.1).
Set the R3765/67G series as follows.
The measurement here is performed with the span lessened and inside of the
bandwidth extended.
Calibrate frequency characteristics. (The same operation as sub-section 3.5.2.)
Note:
OPT11 or OPT14 uses a different [MEAS] menu.
For more information, refer to the description on page 7-10.
Block name
Setting
Key operation
ACTIVE
CHANNEL
Set the channel to 2.
[CH 2]
RESPONSE
Select the input port in the
receiver part.
Set the measurement
format to group delay
display.
AG type :
[
MEAS
]
{B/R}
(Initial setup)
BG type :
[MEAS
]
{TRANSMISSION}
(Initial setup)
CG type :
[MEAS
]
{S21 TRANS FWD} (Initial setup)
[FORMAT
]
{DELAY}
STIMULUS
Center frequency
880MHz
Span frequency
50MHz
[CENTER]
[8]
[8]
[0]
[MHz]
[SPAN]
[5]
[0]
[MHz]
1
2
3