R3765/67G Series Network Analyzer Operation Manual
8.2 Error Messages
8-4
8.2.3
Operating Error
Calibration aborted.
Memorizing calibration data was aborted.
<How to handle>
While calibration data is being memorized, if the setting is
changed, the calibration is aborted.
Do not change the setting until the calibration is finished. (Refer
to sub-section 7.5.5 and 7.5.6.)
Calibration canceled !
The acquired calibration data was cleared because the sweeping
conditions were changed during calibration operated.
<How to handle>
The sweeping conditions must not be changed to acquire more
than two calibration data.
Execute the calibration data acquisition from the first. (Refer to
sub-section 7.5.5.)
Calibration data not
found.
CORRECT ON was executed without memorized calibration
data.
<How to handle>
Memorize the calibration data. (Refer to sub-section 7.5.5.)
Can’t find plotter !!!
A plotter was not found in a plot output.
<How to handle>
The plotter is not connected or GPIB address of the plotter is not
correct. (Refer to sub-section 7.13.2.)
Can’t ... When
CORRECT ON.
To memorize calibration data or to execute CLEAR CAL DATA
was attempted in the state of CORRECT ON.
<How to handle>
Choose CORRECT OFF. (Refer to sub-section 7.5.10.)
Can’t ... When PROG-
SWEEP.
The operator attempted to change the setting of the number of
points while in the program sweep.
<How to handle>
Specify a sweep type other than PROGRAM SWEEP and USER
SWEEP. (Refer to section 7.9.)