R3765/67G Series Network Analyzer Operation Manual
6.3 Saving to the Floppy Disk
6-9
Put the cursor on the file to recall with {CURSOR } and {CURSOR }.
By {LOAD} the display on the screen becomes as follows.
Figure 6-8 Recall from Floppy Disk
After the completion of the recall, the sweep is in the held state automatically.
6.3.1
About the Measurement Data to Save
There are three kinds in the measurement data to save.
(1)
RAW ARRAY (Raw data)
(2)
FORMAT ARRAY (Format data)
(3)
MEM ARRAY (Memory data)
Now (1) RAW ARRAY and (2) DATA ARRAY save the display data.
The difference between the two data is as follows.
•
RAW ARRAY saves the data before the error correction, the trace computation, etc.
processed.
•
DATA ARRAY saves the displayed data itself.
For example, when the saved data is recalled by RAW ARRAY, the right value when saved can
be displayed even if the measurement format is changed.
In DATA ARRAY, if the displayed format when saved is of LOG MAG setting, the right value is
displayed only with LOG MAG format at the data recalling.
Refer to Figure 9-1 for the flow of three data.