R3765/67G Series Network Analyzer Operation Manual
8.2 Error Messages
8-10
Warning of Internal Set, Change, etc.
Display Mode changed.
The display mode setting was internally altered to DISPLYA
DATA.
<How to handle>
In the memory trace display (DISPLAY MEMORY,DISPLAY
DATA&MEM), the measuring condition under which the memory
trace was acquired must be the same as the current measuring
condition and the measuring format must be set to one trace.
Therefore, when the number of points or the sweep type is
altered in a state that the memory trace is displayed, or when the
measuring format is set to two traces (LOGMAG&PHASE,
LOGMAG&DELAY or LINMAG&PHASE), this message is
displayed and the display mode is altered to DISPLAY DATA
internally. (Refer to sub-section 7.4.4.)
Sweep time increased.
The setting of the sweep time was internally altered and the
sweep time was increased.
<How to handle>
The minimum setting value of the sweep time is decided
according to the RBW setting, etc. When the sweep time is set to
AUTO, this message is not displayed. Therefore, when the
sweep time is not set to AUTO, if this message is displayed by
altering the setting of the RBW or and the sweep time is
increased.
Afterward, even if the RBW setting is set to the previous setting,
the sweep time setting do not be set back to the previous setting.
(Refer to sub-section 7.9.3.)
Trace-Math turned off.
The setting of the trace operation (DATA+MEM and others) was
internally altered to OFF.
<How to handle>
The measuring condition in which the memory trace was
obtained must be the same as the current measuring condition in
the trace operation.
Therefore, when the number of points or the sweep type was
altered with the trace operation executed, this message is
displayed and the trace operation is set to OFF. (Refer to sub-
section 7.4.7.)