7.12 Limit Function
7-136
Edit segment menu
{STIMULUS VALUE}
: Sets stimulus value of the segment with ENTRY block.
{MARKER TO STIMULUS}: Sets stimulus value of the segment with active marker.
Turning the data knob moves the active marker right and
left.
{UPPER LIMIT}
: Sets the upper limit value of the segment.
It is necessary to set both values, upper limit and lower
limit.
If the upper limit value is not required, set an extreme
large value for the upper limit value.
Pressing {UPPER LIMIT} or {LOWER LIMIT} key
changes limit values displayed on the screen into upper/
lower expression, even if they are set with middle/delta
values.
If a value smaller than the lower limit is input for the
upper limit value or the reverse, the same values are set
for both the limit values.
{LOWER LIMIT}
: Sets the lower limit value of the segment.
It is necessary to set both values, upper limit and lower
limit.
If the lower limit value is not required, set an extreme
small value for the lower limit value.
{DELTA LIMIT}
: Sets the limit width of the segment.
The limit width is expressed with the center value set by
{MIDDLE VALUE}.
For example, to set the pass area within -5dB
±
3dB,
enter -5dB as the center value and 6dB as the delta
value.
Pressing {MIDDLE LIMIT} or {DELTA LIMIT} key
changes limit values displayed on the screen into upper/
lower expression, even if they are set with DELTA/
MIDDLE values.
{MIDDLE VALUE}
: Sets the middle value of DELTA LIMIT.
{MARKER TO MIDDLE}
: Sets the middle value to the active marker position.
Clear limit menu
{YES}
: Clears the Limit Table and returns to the edit limit menu.
{NO}
: Returns to the edit limit menu without clearing the Limit
Table.
8