R3765/67G Series Network Analyzer Operation Manual
7.5 Calibration
7-56
Press {DONE TRANS} to execute the transmission calibration.
When the transmission calibration has been completed, the 3-port full calibration
menu is displayed.
Note
It is possible to retrieve calibration data for each calibration standard if {DONE
TRANS} has not been pressed yet.
Press {ISOLATION} to call the isolation menu.
To omit the isolation calibration, press {OMIT ISOLATION}, then press {DONE ISOLA-
TION}.:
Follow directions below to execute the isolation calibration:
(25-A)Connect the load standard between TEST PORT1 and TEST PORT2, then press
{P1-P2 ISOLATION}.
A message "Wait for Sweep ." is displayed and the calibration data is acquired.
When the message disappears, the calibration data acquisition is completed.
(Note)
(25-B) Connect the load standard between TEST PORT1 and TEST PORT3, then press
{P1-P3 ISOLATION}.
A message "Wait for Sweep ." is displayed and the calibration data is acquired.
When the message disappears, the calibration data acquisition is completed.
(Note)
(25-C) Connect the load standard between TEST PORT2 and TEST PORT3, then press
{P2-P3 ISOLATION}.
A message "Wait for Sweep ." is displayed and the calibration data is acquired.
When the message disappears, the calibration data acquisition is completed.
(Note)
(25-D)Press {DONE ISOLATION} to execute the isolation calibration.
When the isolation calibration has been completed, the 3-port full calibration menu
is displayed.
Note:
It is possible to retrieve calibration data for each calibration standard if
{DONE
ISOLATION}
has not been pressed yet.
Press {DONE 3-PORT} to execute the 3-port full calibration.
Note: Do not move the instrument, connected cables, connectors, or calibration
standard while the message "Wait for Sweep ." is displayed.
If the set conditions are changed while the message is displayed, the
"Calibration aborted ." message is displayed, and calibration data cannot be
retrieved during that time.