R3765/67G Series Network Analyzer Operation Manual
10.8 Crosstalk
10-36
(5)
Crosstalk of TEST PORT 2 (When the measurement path is S23)
Testing procedure
Setup the R3765CG/67CG as follows.
Figure 10-21 Crosstalk of TEST PORT 2 (When the measurement path is S23)
Setup the R3765CG/67CG.
MEAS
: S23
RBW
: 100Hz
Average : 16 times
Connect the short standard to TEST PORT 3.
Connect the load standard to TEST PORT 2.
Read the crosstalk value of TEST PORT 2 from the trace data.
<Check> :
Crosstalk of TEST PORT 2
OPT 11
300kHz to 40MHz : -90dB or less
40MHz to 2.6GHz : -100dB or less
2.6GHz to 3.8GHz : -90dB or less
3.8GHz to 5.0GHz : -80dB or less (R3767CG only)
5.0GHz to 8.0GHz : -70dB or less (R3767CG only)
OPT 13
300kHz to 40MHz : -85dB or less
40MHz to 3.8GHz : -90dB or less
1
TEST
PORT1
TEST
PORT2
TEST
PORT3
R3765CG/67CG
(OPT11/OPT13/OPT14)
LO
AD
st
and
ard
SHOR
T
st
and
ard
TEST
PORT4
2
3
4
5
6