R3765/67G Series Network Analyzer Operation Manual
7.4 RESPONSE Block
7-29
Depending on the measurement frequency, the IF RBW
AUTO function automatically switches the resolution
bandwidth during sweeping since the resolution
bandwidth cannot be distinguished within the sweep
frequency range.
When [AUTO] is set, the resolution bandwidth is
automatically selected according to the measurement
frequency range as shown in the table below. Then the
measurement is performed.
300Hz
3.4msec/POINT
200Hz
5.0msec/POINT
150Hz
7.0msec/POINT
100Hz
11.0msec/POINT
70Hz
14.0msec/POINT
50Hz
19.0msec/POINT
40Hz
26.1msec/POINT
30Hz
34.9msec/POINT
20Hz
50.1msec/POINT
15Hz
70.1msec/POINT
10Hz
99.3msec/POINT
Measurement frequency
Resolution bandwidth
300kHz to 450kHz
10kHz
450kHz to 700kHz
15kHz
700kHz to 8GHz
20kHz
Resolution bandwidth
Maximum sweeping per point