R3765/67G Series Network Analyzer Operation Manual
7.5 Calibration
7-40
Connect the short standard to port 2 (inverse direction reflection measurement port)
and press {S22 (PORT2) REV:SHORT}.
The message “Wait for Sweep .” is displayed and the calibration data is obtained. The
calibration is completed when the message disappears.
Connect the load standard to port 2 (inverse direction reflection measurement port)
and press {S22 (PORT2) REV:LOAD}.
The message “Wait for Sweep .” is displayed and the calibration data is obtained. The
calibration is completed when the message disappears.
Pressing {DONE REFLECT’N} executes the reflection calibration.
When the reflection calibration is completed, returns to the 2-port full calibration menu.
CAUTION!
The calibration data of each calibration standard can be acquired again before {DONE RE-
FLECT’N} pressed.
Press {TRANSMISSION} to call the 2-port transmission menu.
Connect the through standard between port 1 and port 2.
Press {GROUP THRU}.
The message “Wait for Sweep .” is displayed and the calibration data is obtained. The
calibration is completed when the message disappears. (Note)
When this operation is performed, the following operation of step
to
is not neces-
sary.
Press {FWD.TRANS THRU}.
The message “Wait for Sweep .” is displayed and the calibration data is obtained. The
calibration is completed when the message disappears. (Note)
Press {FWD.MATCH THRU}.
The message “Wait for Sweep .” is displayed and the calibration data is obtained. The
calibration is completed when the message disappears. (Note)
Press {REV.TRANS THRU}.
The message “Wait for Sweep .” is displayed and the calibration data is obtained. The
calibration is completed when the message disappears. (Note)
Press {REV.MATCH THRU}.
The message “Wait for Sweep .” is displayed and the calibration data is obtained. The
calibration is completed when the message disappears. (Note)
Pressing {DONE TRANS} executes transmission calibration.
Returns to 2-port calibration menu when the transmission calibration is completed.
CAUTION!
The calibration data of each calibration standard can be acquired again before {DONE
TRANS} pressed.
Press {ISOLATION} to call the 2-port isolation menu.
11
12
13
14
15
16
16
19
17
18
19
20
21
22