R3765/67G Series Network Analyzer Operation Manual
5.8 GO/NG Measurement Using Limit Line Function
5-47
•
Establish the edit mode.
[SYSTEM]
{LIMIT MENU}
{EDIT LIMIT LINE}
The display on the screen is as follows.
Figure 5-40 Limit Line Editing
•
Set each segment.
Setting of segment 0.
SEGMENT 0
: {EDIT SEGMENT}
{STIMULUS VALUE}
[8]
[3]
[0]
[MHz]
{UPPER LIMIT}
[-]
[4]
[0]
[X1]
{LOWER LIMIT}
[-]
[6]
[5]
[X1]
{Return}
As the marker can be used with data knob, etc. now, it’s useful to confirm the setting
value of each segment.